Relative Humidity
 | 
| Sensor | 
HUMICAP 180R | 
| Measurement Range | 
0 to 100% RH (non-condensing) | 
| Response Time | 
- The response time for the RH specification is for the HUMICAP 180R at 20°C in still air with sintered PTFE filter and a 0 to 75% RH step change.
 
- 20 s (63% step change)
 
- 60 s (90% step change)
 
 
 | 
| Factory Calibration Uncertainty | 
- The factory calibration uncertainty is defined as ±2 standard deviation limits. Uncertainty is at +20°C. Small variations are possible; see also the calibration certificate.
 
- ±0.6% RH 0 to 40% RH
 
- ±1.0% RH 40 to 97% RH
 
 
 | 
| -NOTE- | 
Accuracy specifications include non-linearity, hysteresis, and repeatability. | 
| Accuracy at 15° to 25°C | 
- ±1% RH (0 to 90% RH)
 
- ±1.7% RH (90 to 100% RH)
 
 
 | 
| Accuracy at -60° to -40°C | 
±(1.4 + 0.032 × reading) % RH | 
| Accuracy at -40° to -20°C | 
±(1.2 + 0.012 × reading) % RH | 
| Accuracy at -20° to +40°C | 
±(1.0 + 0.008 × reading) % RH | 
| Accuracy at 40° to 60°C | 
± (1.2 + 0.012 × reading) % RH |